2026

Mismatch-Driven CRISPR/Cas12a Biosensing of UV-Induced DNA Lesions for Environmental Solar Exposure Surveillance.

Environmental science & technology Chen YW, Marpaung DSS, Chen YY, Singuru MMR, Chuang MC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Monitoring environmentally relevant ultraviolet (UV) radiation is critical for understanding its biological impacts on ecosystems and human health. However, conventional UV dosimeters lack the molecular sensitivity to detect DNA-level damage that initiates such effects. Here, we present a CRISPR/Cas12a-based biosensing platform capable of quantifying solar UV exposure through the detection of UV-induced thymine dimers in DNA activators. This system harnesses mismatch-driven suppression of Cas12a activity, enabling a reduction in the fluorescence signal in response to UV-induced molecular lesions. The impact of thymine arrangement and the dimerization position of the activators on sensitivity were investigated. UV-induced diminution in Cas12a's trans-cleavage efficiency (kcat/Km) was also characterized, revealing a 1.67-fold decrease as the UVB dose increased from 0 to 2 J/cm2. Under optimized conditions, the sensor achieved a detection limit of 0.029 J/cm2 for UVB and demonstrated high sensitivity to UVC. Field validation under natural sunlight showed a strong correlation with reference radiometric measurements, validating the biosensor's accuracy and environmental relevance. The system's sensitivity to low lesion densities, straightforward mechanism, and simple operation highlights its potential for environmental surveillance, human health risk assessment, and ecological monitoring in response to solar UV radiation.

关键词