Graphene-Based Chemical Field-Effect Transistors: Impact of Electric Double Layer Model and Quantum Capacitance on Na

Micromachines Baridi G, Liaquat A, Martini L, Nappi L, Rapuzzi F, Clericò V, Abidi EH, Meziani YM, Amado M, Diez E, Brancolini G, Rovati L, Rossella F
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Graphene-based ion-sensitive field-effect transistors can operate as biosensors by utilizing the formation of an electric double layer at the interface between the electrolyte and the graphene channel, enabling high sensitivity, scalability, and cost-effective fabrication. In this work, we focus on the working principles and current methodologies associated with these devices, making a comparative analysis of different models that describe the electric double layer in the electrolyte, referring to sodium ions (Na+) as a case study for the detection performance of the graphene biosensor, and taking into account the impact of graphene quantum capacitance. Our study addresses the sensitivity of graphene field-effect transistors within the framework of the Gouy-Chapman model, as well as the Stern model, computing device sensitivities of 3200 V/M and 5500 V/M, respectively. By incorporating the impact of graphene's quantum capacitance in the calculations, increased sensitivity up to 5620 V/M was found. The present work shines light on the rationalization of graphene-based biosensors' operation and performance.