2026

Low-Effort Respiratory Function Estimation with a Soft Wearable Digital Spirometry Patch.

Biosensors Karim FA, Tariq A, Fitzpatrick CB, Zhou L, Suárez-Fariñas M, Schotland H, Rogers L, Lee YJ, Yeo WH, Kim YS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Spirometry is widely regarded as the clinical gold standard for quantifying lung function. It plays a central role in the diagnosis and management of cardiopulmonary disorders, including asthma and chronic obstructive pulmonary disease (COPD). However, the procedure relies on a forceful and often stressful expiratory maneuver that may cause patient discomfort and require substantial effort, frequently necessitating active coaching and trained personnel to ensure reproducible measurements. In this paper, we present the Digital Spirometry Patch (DSP), a soft, flexible, wearable patch capable of estimating lung function parameters by utilizing low-effort breathing maneuvers. Eighteen participants performed low-effort and forceful breathing maneuvers while wearing the DSP to collect tracheal sound and chest movement signals for spirometric parameter estimation using elastic net and simple linear regression. Using leave-one-subject-out cross-validation, the elastic net models achieved RMSEs of 0.668 L, 0.224 L, and 0.428 L/s for FVC, FEV1, and PEF, respectively, using low-effort breathing maneuvers, and 0.499 L, 0.304 L, and 0.891 L/s using forceful exhalation maneuvers. These results demonstrate the potential of the DSP as a wearable, low-effort alternative for estimating lung function outside of conventional spirometry settings.

关键词