组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Spirometry is widely regarded as the clinical gold standard for quantifying lung function. It plays a central role in the diagnosis and management of cardiopulmonary disorders, including asthma and chronic obstructive pulmonary disease (COPD). However, the procedure relies on a forceful and often stressful expiratory maneuver that may cause patient discomfort and require substantial effort, frequently necessitating active coaching and trained personnel to ensure reproducible measurements. In this paper, we present the Digital Spirometry Patch (DSP), a soft, flexible, wearable patch capable of estimating lung function parameters by utilizing low-effort breathing maneuvers. Eighteen participants performed low-effort and forceful breathing maneuvers while wearing the DSP to collect tracheal sound and chest movement signals for spirometric parameter estimation using elastic net and simple linear regression. Using leave-one-subject-out cross-validation, the elastic net models achieved RMSEs of 0.668 L, 0.224 L, and 0.428 L/s for FVC, FEV1, and PEF, respectively, using low-effort breathing maneuvers, and 0.499 L, 0.304 L, and 0.891 L/s using forceful exhalation maneuvers. These results demonstrate the potential of the DSP as a wearable, low-effort alternative for estimating lung function outside of conventional spirometry settings.