2026

A robust fluorescent glycerol kinase biosensor for reliable quantification of glycerol.

Talanta Sorenson AE, Schaeffer PM
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Glycerol is a natural component of many biological samples, beverages and foodstuffs making accurate quantification relevant to a wide variety of industries and applications. Here we present a simple, rapid, and reliable biosensor for quantification of glycerol in a variety of sample types including water, alcoholic beverages, balsamic vinegar, honey, syrup and artificial human urine. This system leverages GFP-tagged Escherichia coli glycerol kinase with differential scanning fluorimetry of GFP-tagged proteins for direct quantification of glycerol via binding-induced increases in thermal stability. The GFP tag enables additional indication of potential matrix effects. The biosensor has a broad linear range of 31 μM - 7.5 mM and a limit of detection of 19 μM. Good agreement was obtained between the fluorescent glycerol kinase biosensor and a commercial glycerol quantification kit with glycerol-spiked artificial urine as well as a variety of alcoholic beverages and foodstuffs.

关键词