组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Background: Non-ST segment elevation acute coronary syndrome (NSTE-ACS) is conventionally diagnosed using electrocardiography and serial blood biomarker measurements. We investigated a noninvasive, bloodless, and electrode-free diagnostic strategy using a wrist-worn infrared spectrophotometric biosensor (Infrasensor). Results: In a prospective multicenter study of 595 patients with suspected NSTE-ACS enrolled across 13 sites in two countries, participants were stratified into five analytical cohorts. With 200 multi-ethnic controls and a leave-one-cohort-out external validation, a machine learning model detected high-grade coronary obstruction with an area under the receiver operating characteristic curve (AUC) of 0.87 (95% CI: 0.84-0.90), 90% specificity, and 84% positive predictive value-surpassing standard risk scores. A secondary model predicted freedom from NSTE-ACS and adverse outcomes over 30 days with an AUC of 0.89 (95% CI: 0.87-0.92), 99% sensitivity, and 96% negative predictive value. Conclusions: The Infrasensor demonstrated high diagnostic accuracy for high-grade coronary obstruction and 30-day adverse outcome prediction, surpassing conventional risk scores across a prospective, multi-ethnic, multicenter cohort. These findings support its potential as a rapid, noninvasive point-of-care tool for early NSTE-ACS risk stratification.