1997

Reagentless mediated laccase electrode for the detection of enzyme modulators.

Analytical chemistry Trudeau F, Daigle F, Leech D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We have investigated aerobic mediation of electron transfer to a laccase enzyme by the solution redox couples [Os(bpy)(2)Cl(2)](+/0) and [Os(bpy)(2)(MeIm)Cl](2+/+), where bpy is 2,2-bipyridine and MeIm is N-methylimidazole. The factors that influence the homogeneous mediation reaction are investigated and discussed. Investigation of ionic strength, pH, and temperature effects on the kinetics of intermolecular electron transfer elucidates the governing factors in the mediator-enzyme reactions. Coimmobilization of both enzyme and an osmium redox mediator in a hydrogel on glassy carbon electrodes results in a biosensor for the reagentless addressing of enzyme activity, consuming only oxygen present in solution. Thus, these immobilized enzyme biosensors can be utilized for the detection of modulators of laccase activity, such as the inhibitor sodium azide. The enzyme inhibition biosensor can detect levels of azide as low as 2.5 × 10(-6) mol dm(-3) in solution.