1999

Amperometric thick-film strip electrodes for monitoring organophosphate nerve agents based on immobilized organophosphorus hydrolase.

Analytical chemistry Mulchandani A, Mulchandani P, Chen W, Wang J, Chen L
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An amperometric biosensor based on the immobilization of organophosphorus hydrolase (OPH) onto screen-printed carbon electrodes is shown useful for the rapid, sensitive, and low-cost detection of organophosphate (OP) nerve agents. The sensor relies upon the sensitive and rapid anodic detection of the enzymatically generated p-nitrophenol product at the OPH/Nafion layer immobilized onto the thick-film electrode in the presence of the OP substrate. The amperometric signals are linearly proportional to the concentration of the hydrolyzed paraoxon and methyl parathion substrates up to 40 and 5 μM, showing detection limits of 9 × 10(-)(8) and 7 × 10(-)(8) M, respectively. Such detection limits are substantially lower compared to the (2-5) × 10(-)(6) M values reported for OPH-based potentiometric and fiber-optic devices. The high sensitivity is coupled to a faster and simplified operation, and the sensor manifests a selective response compared to analogous enzyme inhibition biosensors. The applicability to river water sampling is illustrated. The attractive performance and greatly simplified operation holds great promise for on-site monitoring of OP pesticides.