2013

Fabrication of an amperometric D-amino acid biosensor based on nickel hexacyanoferrate polypyrrole hybrid film deposited on glassy carbon electrode.

Bioprocess and biosystems engineering Lata S, Pundir CS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A nickel hexacyanoferrate polypyrrole film was synthesized through an electrochemical two-step methodology leading to a very stable and homogenous robust hybrid film. A highly sensitive, specific and rapid amperometric D-amino acid biosensor was constructed by immobilizing D-amino acid oxidase on this film deposited over the surface of a glassy carbon electrode. The modified electrode was characterized by scanning electron microscopy, electrochemical impedance spectroscopy and Fourier transform infrared spectrophotometry. The biosensor showed optimum response within 1 s, when operated at 50 mV s(-1) in 0.01 M Tris HCl buffer, pH 7.0 at 30 °C. The biosensor exhibited excellent sensitivity with a detection limit of 1.5 µM (S/N = 3) for D-amino acids and wider linear range, 20-500 µM. Analytical recovery of added D-alanine (5 and 10 mM) in serum samples was 98.00 and 98.80 %, respectively. Within-batch and between-batch coefficients of variation in serum samples were 1.36 and 2.77 %, respectively. The enzyme electrode was used more than 50 times over 2 months, when stored at 4 °C. The proposed modified electrode exhibited sufficient mechanical and electrochemical stability and high sensitivity compared to earlier electrochemical D-amino acid biosensors. Interference by ascorbic acid and uric acid, the main interfering species in the biological samples, was negligible.

关键词