2013

An Fe3O4-nanoparticles-based amperometric biosensor for creatine determination.

Artificial cells, nanomedicine, and biotechnology Kaçar C, Erden PE, Pekyardimci S, Kiliç E
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

An amperometric biosensor for the detection of creatine was designed, based on carbon paste electrode modified with Fe(3)O(4) nanoparticles. Electron transfer properties of unmodified and Fe(3)O(4)-nanoparticles-modified carbon paste electrodes were investigated by cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) methods. Fe(3)O(4) nanoparticles increased the surface area and electric conductivity of the electrode, thus enhancing the sensitivity of the electrode. Optimum pH, buffer concentration, working potential and enzyme loading were selected as 7.0; 0.05 mol L(-1); +0.30 V and 2.0 Unit creatinase (CI), 1.0 Unit sarcosine oxidase (SO), respectively. The purposed biosensor exhibited linear response from 2.0 × 10(-7) mol L(-1) to 3.8 × 10(-6) mol L(-1) and from 9.0 × 10(-6) mol L(-1) to 1.2 × 10(-4) mol L(-1) with a detection limit of 2.0 × 10(-7) mol L(-1). Biosensor was used for determination of creatine in commercial creatine powder samples and showed a good sensing performance.

关键词