Construction of an amperometric TG biosensor based on AuPPy nanocomposite and poly (indole-5-carboxylic acid) modified Au electrode.

Bioprocess and biosystems engineering Narang J, Chauhan N, Rani P, Pundir CS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A method is described for construction of an amperometric triglyceride (TG) biosensor based on covalent co-immobilization of lipase, glycerol kinase and glycerol-3-phosphate oxidase onto gold polypyrrole nanocomposite decorated poly indole-5-carboxylic acid electrodeposited on the surface of a gold electrode. The enzyme electrode was characterized by transmission electron microscopy, scanning electron microscopy, electrochemical impedance studies, Fourier transform infrared spectroscopy and cyclic voltammetry. Biosensor showed optimum response within 4 s at pH 6.5 and 35 °C, when polarized at +0.1 V against Ag/AgCl. There was a linear relationship between sensor response and triolein concentration in the range 50-700 mg/dl. Biosensor was employed for determination of TG in serum. Detection limit of the biosensor was 20 mg/dl. Biosensor was evaluated with 91-95 % recovery of added triolein in sera and 4.14 and 5.85 % within and between batch coefficients of variation, respectively. There was a good correlation (r = 0.99) between sera TG values by standard method (Enzymic colorimetric) and the present method. The biosensor was unaffected by a number of serum substances at their physiological concentration. Biosensor lost 50 % of its initial activity after its 100 uses over 7 months, when stored at 4 °C.