组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
The present work aims to improve the sensitivity of an electrical biosensor by simply changing a surface property of the passivation layer, which covers the background region except for the sensing site for electrical isolation among adjacent interconnection lines. The hydrophobic passivation layer dramatically enhances the sensitivity of the biosensor when compared with a hydrophilic passivation layer. A revamped metal oxide semiconductor field-effect transistor (MOSFET), which has a designed underlap region between a gate and a drain, is used as the electrical biosensor. A thin film of CYTOP(TM) and silicon nitride is used as the hydrophobic and hydrophilic passivation layers, respectively. The surface antigen and its specific antibody of the avian influenza virus were employed as the probe and target biomolecule, respectively, to confirm the enhanced sensitivity of the proposed biosensor. By using hydrophobic passivation, the limit of detection of the biosensor was improved up to 100-fold compared with that resulting from hydrophilic passivation. Therefore, a simple surface engineering to control surface wettability can notably improve the sensitivity of a biosensor without additional efforts, such as modifying the sensor structure, optimizing the bio-treatment protocol, or increasing the binding yield between a probe and its target, among other efforts.