2013

CRP detection from serum for chip-based point-of-care testing system.

Biosensors & bioelectronics Kim CH, Ahn JH, Kim JY, Choi JM, Lim KC, Park TJ, Heo NS, Lee HG, Kim JW, Choi YK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Most of point-of-care testing (POCT) to improve facilitates in diagnosis, treatment, and monitoring of patients. POCT technique has still remained a quantitatively and accurately detective effect. In this article, we demonstrated that real human C-reactive protein (CRP) in serum was detected for a chip-based point-of-care testing application based on a nanogap-embedded field effect transistor (FET), and the results were compared with those obtained via the enzyme-linked immunosorbent assay (ELISA) method. The limit of detection (LOD), determined from the standard curve, was 0.1 ng/ml, which is comparable to that of commercialized ELISAs. We evaluated that an improved detection range (0.1 ng/ml to 100 ng/ml) was achieved by comparing with commercialized ELISA. Control experiments to determine selectivity and to discern false-positive/false-negative rates were also performed. This report is the first description of the detection of CRP in human serum using a silicon-based biosensor.

关键词