2013

p-Benzoquinone-mediated amperometric biosensor developed with Psychrobacter sp. for toxicity testing of heavy metals.

Biosensors & bioelectronics Wang X, Liu M, Wang X, Wu Z, Yang L, Xia S, Chen L, Zhao J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A rapid and reliable p-benzoquinone-mediated amperometric biosensor (ToxTell) incorporated with Psychrobacter sp. to detect toxicities of heavy metal ions has been developed. This ToxTell biosensor relied on the real-time monitoring of inhibition effect for metabolism by toxicant to provide early detection and assessment of the degree of toxicity to living cells. The effect of growth phase on the sensitivity of Psychrobacter sp. biosensor was studied. The results showed that at the middle of the logarithmic phase or transition from logarithmic to stationary phase, the Psychrobacter sp. ToxTell biosensor had a higher sensitivity to toxicants. The effects of pH, salinity in respiratory substrates and incubation time on the performance of Psychrobacter sp. biosensor were also investigated. EC(50) values of Cu(2+), Cd(2+), Zn(2+), Cr(6+), Hg(2+) and Pb(2+) to Psychrobacter sp. determined at incubation time 30 min were 2.6 mg/L, 47.3 mg/L, 10.9 mg/L, 14.0 mg/L, 0.8 mg/L and 110.1 mg/L, respectively. The ToxTell microbial biosensor developed in this work demonstrated excellent storage stability for more than 60 days. The biosensor could incorporate different microbial species as biocomponent to reflect the comprehensive values for toxicants in real samples and the results therefore had high degree of validity.

关键词