2013

Development of a fast and sensitive glucose biosensor using iridium complex-doped electrospun optical fibrous membrane.

Analytical chemistry Zhou C, Shi Y, Ding X, Li M, Luo J, Lu Z, Xiao D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Polystyrene electrospun optical fibrous membrane (EOF) was fabricated using a one-step electrospinning technique, functionalized with glucose oxidases (GOD/EOF), and used as a quick and highly sensitive optical biosensor. Because of the doped iridium complex, the fibrous membrane emitted yellow luminescence (562 nm) when excited at 405 nm. Its luminescence was significantly enhanced with the presence of extremely low concentration glucose. The detection limit was of 1.0 × 10(-10) M (S/N = 3), superior to that of reported glucose biosensor with 1.2 × 10(-10) M. A linear range between the relative intensity increase and the logarithm of glucose concentration was exhibited from 3.0 × 10(-10) M to 1.3 × 10(-4) M, which was much wider than reported results. Notably, the response time was less than 1 s. These high sensitivity and fast response were attributed to the high surface-area-to-volume of the porous fibrous membrane, the efficient GOD biocatalyst reaction on the fibers surface, as well as the fast electron or energy transfer between dissolved oxygen and the optical fibrous membrane.

关键词