A highly sensitive reverse sandwich immunoassay for the detection of human cardiac myoglobin (cMb) in serum was designed utilizing a gold nanoparticle (AuNP)-enhanced surface plasmon resonance (SPR) biosensor. First, a monoclonal anti-cMb antibody (Mab1) was covalently immobilized on the sensor surface. AuNPs were covalently conjugated to the second monoclonal anti-cMb antibody (Mab2) to form an immuno-gold reagent (Mab2-AuNP). The reverse sandwich immunoassay consists of two steps: (1) mixing the serum sample with Mab2-AuNP and incubation for the formation of cMb/Mab2-AuNP complexes and (2) sample injection over the sensor surface and evaluation of the Mab1/cMb/Mab2-AuNP complex formation, with the subsequent calculation of the cMb concentration in the serum. The biosensor signal was amplified approximately 30-fold compared with the direct reaction of cMb with Mab1 on the sensor surface. The limit of detection of cMb in a human blood serum sample was found to be as low as 10 pM (approx. 0.18 ng mL(-1)), and the inter-assay coefficient of variation was less than 3%. Thus, the developed SPR-based reverse sandwich immunoassay has a sensitivity that is sufficient to measure cMb across a wide range of normal and pathological concentrations, allowing an adequate estimation of the disease severity and the monitoring of treatment.