2013

Electrocatalytic assay of mercury(II) ions using a bifunctional oligonucleotide signal probe.

Analytica chimica acta Zhang Z, Yin J, Wu Z, Yu R
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Engineered nucleic acid probes containing recognition and signaling functions find growing interest in biosensor design. In this paper, we developed a novel electrochemical biosensor for sensitive and selective detecting of Hg(2+) based on a bifunctional oligonucleotide signal probe combining a mercury-specific sequence and a G-quadruplex (G4) sequence. For constructing the electrochemical Hg(2+) biosensor, a thiolated, mercury-specific oligonucleotide capture probe was first immobilized on gold electrode surface. In the presence of Hg(2+), a bifunctional oligonucleotide signal probe was hybridized with the immobilized capture probe through thymine-mercury(II)-thymine interaction-mediated surface hybridization. The further interaction between G4 sequence of the signal probe and hemin generated a G4-hemin complex, which catalyzed the electrochemical reduction of hydrogen peroxide, producing amplified readout signals for Hg(2+) interaction events. This electrochemical Hg(2+) biosensor was highly sensitive and selective to Hg(2+) in the concentration of 1.0 nM to 1 μM with a detection limit of 0.5 nM. The new design of bifunctional oligonucleotide signal probes also provides a potential alternative for developing simple and effective electrochemical biosensors capable of detecting other metal ions specific to natural or artificial bases.

关键词