2013

Development of an amperometric screen-printed galactose biosensor for serum analysis.

Analytical biochemistry Kanyong P, Pemberton RM, Jackson SK, Hart JP
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The development of a disposable amperometric biosensor for the measurement of circulating galactose in serum is described. The biosensor comprises a screen-printed carbon electrode (SPCE), incorporating the electrocatalyst cobalt phthalocyanine (CoPC), which is covered by a permselective cellulose acetate (CA) membrane and a layer of immobilized galactose oxidase (GALOX). The optimal response of the biosensor, designated as GALOX-CA-CoPC-SPCE, was obtained by systematically examining the effects of enzyme loading, temperature, pH, and buffer strength. The optimal performance of the biosensor occurred with 2U of GALOX, at 35°C, using 50mM phosphate buffer solution (pH 7.0). The sensitivity was 7.00μAmM(-1)cm(-2) and the linear range from 0.1 to 25mM with a calculated limit of detection (LOD) of 0.02mM; this concentration range and LOD are appropriate to diagnose galactosemia, i.e., concentrations >1.1mM in infants. When the biosensor was used in conjunction with amperometry in stirred solution for the analysis of serum, the precision values obtained on unspiked (endogenous level of 0.153mM) and spiked serum (1mM added) (n=6) were 1.10% and 0.11%, respectively, with a calculated recovery of 99.9%.

关键词