Quantification of ovarian cancer markers with integrated microfluidic concentration gradient and imaging nanohole surface plasmon resonance.

The Analyst Escobedo C, Chou YW, Rahman M, Duan X, Gordon R, Sinton D, Brolo AG, Ferreira J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Nanohole array-based biosensors integrated with a microfluidic concentration gradient generator were used for imaging detection and quantification of ovarian cancer markers. Calibration curves based on controlled concentrations of the analyte were created using a microfluidic stepped diffusive mixing scheme. Quantification of samples with unknown concentration of analyte was achieved by image-intensity comparison with the calibration curves. The biosensors were first used to detect the immobilization of ovarian cancer marker antibodies, and subsequently applied for the quantification of the ovarian cancer marker r-PAX8 (with a limit of detection of about 5 nM and a dynamic range from 0.25 to 9.0 μg.mL(-1)). The proposed biosensor demonstrated the ability of self-generating calibration curves on-chip in an integrated microfluidic platform, representing a further step towards the development of comprehensive lab-on-chip biomedical diagnostics based on nanohole array technology.