2013

Label-free detection of cardiac troponin-I using carbon nanofiber based nanoelectrode arrays.

Analytical chemistry Periyakaruppan A, Gandhiraman RP, Meyyappan M, Koehne JE
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A label-free biosensor is presented using carbon nanofiber (CNF) nanoelectrode arrays for the detection of cardiac troponin-I in the early diagnosis of myocardial infarction. Immobilization of anti-cTnI Ab on CNFs and the detection of human-cTnI were examined using electrochemical impedance spectroscopy and cyclic voltammetry techniques. Each step of the modification process was monitored, and the results show changes in electrical capacitance or resistance to charge transfer due to the specificity of corresponding adsorption of Ab-Ag interaction. The immunosensor demonstrates a good selectivity and high sensitivity against human-cTnI analytes and is capable of detecting cTnI at concentrations as low as ∼0.2 ng/mL, which is 25 times lower than that possible by conventional methods. Analysis of the electrode at various stages using atomic force microscopy and X-ray reflectivity provides information on the surface roughness and orientation of the antibody.

关键词