2013

Development of a biosensor for detection of pleural mesothelioma cancer biomarker using surface imprinting.

PloS one Mathur A, Blais S, Goparaju CM, Neubert T, Pass H, Levon K
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Hyaluronan-linked protein 1 (HAPLN1) which has been shown to be highly expressed in malignant pleural mesotheliomas (MPM), was detected in serum using an electrochemical surface-imprinting method. First, the detection method was optimized using Bovine serum albumin (BSA) as a model protein to mimic the optimal conditions required to imprint the similar molecular weight protein HAPLN1. BSA was imprinted on the gold electrode with hydroxyl terminated alkane thiols, which formed a self-assembled monolayer (SAM) around BSA. The analyte (BSA) was then washed away and its imprint (empty cavity with shape-memory) was used for detection of BSA in a solution, using electrochemical open-circuit potential method, namely potentiometry. Factors considered to optimize the conditions include incubation time, protein concentration, limit of detection and size of electrode. Matrix assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) was used to confirm selectivity of imprints. With the obtained imprinting control parameters, HAPLN1 was imprinted in duplicate and the detection of spiked HAPLN1 was successfully conducted in serum.

关键词