2013

Small organic molecules detection based on aptamer-modified gold nanoparticles-enhanced quartz crystal microbalance with dissipation biosensor.

Analytical biochemistry Zheng B, Cheng S, Liu W, Lam MH, Liang H
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Small molecules are difficult to detect by the conventional quartz crystal microbalance with dissipation (QCM-D) technique directly because the changes in frequency resulting from the binding processes of small biomolecules are often small. In the current study, an aptamer-based gold nanoparticles (AuNPs)-enhanced sensing strategy for detection of small molecules was developed. The QCM crystal was first modified with a layer of thiolated linker DNA, which can be partly base-paired with the detection part containing the adenosine aptamer sequence. In the presence of adenosine, the aptamer bound with adenosine and folded to the complex structure, which precluded the reporter part carrying AuNPs to combine with the random coiled detection part. Therefore, the lower the concentration of adenosine, the more AuNPs combined to the crystal. The resulting aptasensor showed a linear response to the increase of the adenosine concentration in the range of 0-2 μM with a linear correlation of r=0.99148 and a detection limit of 65 nM. Moreover, the aptasensor exhibited several excellent characteristics such as high sensitivity, selectivity, good stability, and reproducibility.

关键词