Optimization of hydrogen peroxide detection for a methyl mercaptan biosensor.

Sensors (Basel, Switzerland) Li ZH, Guedri H, Viguier B, Sun SG, Marty JL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Several kinds of modified carbon screen printed electrodes (CSPEs) for amperometric detection of hydrogen peroxide (H₂O₂) are presented in order to propose a methyl mercaptan (MM) biosensor. Unmodified, carbon nanotubes (CNTs), cobalt phthalocyanine (CoPC), Prussian blue (PB), and Os-wired HRP modified CSPE sensors were fabricated and tested to detect H₂O₂, applying a potential of +0.6 V, +0.6 V, +0.4 V, -0.2 V and -0.1 V (versus Ag/AgCl), respectively. The limits of detection of these electrodes for H₂O₂ were 3.1 μM, 1.3 μM, 71 nM, 1.3 μM, 13.7 nM, respectively. The results demonstrated that the Os-wired HRP modified CSPEs gives the lowest limit of detection (LOD) for H₂O₂ at a working potential as low as -0.1 V. Os-wired HRP is the optimum choice for establishment of a MM biosensor and gives a detection limit of 0.5 μM.

关键词