2013

Measurement of porcine haptoglobin in meat juice using surface acoustic wave biosensor technology.

Meat science Klauke TN, Gronewold TM, Perpeet M, Plattes S, Petersen B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Aim of the study was the application of biosensor technique to measure the concentration of an acute phase protein (APP) within complex matrices from animal origin. For the first time, acute phase protein haptoglobin (Hp) was detected from unpurified meat juice of slaughter pigs by a label-free biosensor-system, the SAW-based sam®5 system. The system uses a sensor chip with specific antibodies to catch Hp while the mass-related phase shift is measured. The concentration is calculated as a function of these measured phase shifts. The results correlate very well with reference measurement results obtained by enzyme-linked immunosorbent assay (ELISA), R=0.98. The robust setup of the surface acoustic wave (SAW)-based system and its ability to measure within very short time periods qualifies it for large-scale analyses and is apt to identify rapidly pigs in the meat production process whose consumption would have an increased risk for consumers.

关键词