2013

Sensitive D-amino acid biosensor based on oxidase/peroxidase system mediated by pentacyanoferrate-bound polymer.

Biosensors & bioelectronics Nieh CH, Kitazumi Y, Shirai O, Kano K
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A sensitive d-amino acid oxidase (DAAO)/peroxidase (POD) bienzyme biosensor is constructed, in which pentacyanoferrate-bound poly(1-vinylimidazole) polymer (PVI[Fe(CN)5]) is selected as a mediator. Reductive current of PVI[Fe(CN)5] related to the H2O2 concentration generated in the DAAO reaction was measured at -0.1V vs. Ag|AgCl with DAAO/POD/PVI[Fe(CN)5]-modified electrode. The result revealed that PVI[Fe(CN)5] is suitable as a mediator for this bienzyme system due to its appropriate formal potential and its extremely low reactivity against DAAO. The stability of DAAO was improved by adding free flavin adenine dinucleotide and the electrode composition was optimized for the detection of d-alanine. Nafion and ascorbate oxidase-immobilized films worked successfully to prevent severe interference from uric acid and ascorbic acid. The low detection limits of d-alanine (2μM) and d-serine (2μM) imply its possibility for the determination of extremely low concentration of d-amino acids in physiological fluids. The proposed bienzyme biosensor is proved to be capable of detecting d-amino acids in urine.

关键词