2013

Disposable biosensor based on platinum nanoparticles-reduced graphene oxide-laccase biocomposite for the determination of total polyphenolic content.

Talanta Eremia SA, Vasilescu I, Radoi A, Litescu SC, Radu GL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A disposable amperometric biosensor was developed for the detection of total polyphenolic compounds from tea infusions. The biosensor was designed by modifying the surface of a carbon screen-printed electrode with platinum nanoparticles and reduced graphene oxide, followed by the laccase drop-casting and stabilization in neutralised 1% Nafion solution. The obtained biosensor was investigated by scanning electron microscopy and electrochemical techniques. It was observed that platinum nanoparticles-reduced graphene oxide composite had synergistic effects on the electron transfer and increased the electroactive surface area of the carbon screen-printed electrode. The constructed analytical tool showed a good linearity in the range 0.2-2 μM for caffeic acid and a limit of detection of 0.09 μM. The value of Michaelis-Menten apparent constant was calculated from the electrochemical version of Lineweaver-Burk equation to be 2.75 μM. This disposable laccase biosensor could be a valuable tool for the estimation of total polyphenolic content from tea infusions.

关键词