2013

Screen-printed biosensor based on the inhibition of the acetylcholinesterase activity for the determination of codeine.

Talanta Asturias-Arribas L, Alonso-Lomillo MA, Domínguez-Renedo O, Arcos-Martínez MJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The current paper presents the chronoamperometric determination of codeine using screen-printed carbon electrodes that incorporate tetrathiafulvalene in the matrix of the working electrode, as mediator, and cross-linked acetylcholinesterase. Applying a potential of +250 mV, a 1mM solution of acetylthiocholine in electrolyte solution pH 7 gives an oxidation signal due to the dimerization of its metabolite after the reaction with the enzyme. This electrochemical signal is decreased by consecutive additions of a solution of codeine, which allows the performance of curves of calibration for the validation of this electrochemical method, giving a reproducibility of 3.31% (n=6) and a capability of detection of 20 µM. This type of inhibition has been studied by means of a Lineweaver-Burk plot. Additionally, the developed biosensor was used for the determination of the quantity of codeine in pharmaceutical commercial tablets and urine samples.

关键词