组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
We present a portable and easy-to-use biosensor platform, allowing for label-free detection of diagnostic markers in undiluted animal serum. Exemplarily, this is shown for the detection of anti-Salmonella antibodies. 1-lambda-Reflectometry was used as detection method, making the new biosensor platform portable, cheap, and robust. As recognition elements, lipopolysaccharides (LPSs) from Salmonella typhimurium bacteria were immobilized as sensitive layer on the transducer to carry out serological tests via a direct assay format. For this purpose, a new surface preparation protocol has been worked out allowing for immobilization of the LPS via hydrophobic interactions. It has been shown that results obtained by 1-lambda-Reflectometry are equivalent to those obtained by the non-portable Reflectometric Interference Spectroscopy setup. The new sensor platform was calibrated in both matrices, buffer and undiluted serum. Good sensitivity, selectivity and intra chip reproducibility have been observed. Furthermore, inter chip reproducibility was examined and recovery rates were found to be between 99 and 117% in undiluted serum.