2013

Single embryo-coupled gate field effect transistor for elective single embryo transfer.

Analytical chemistry Sakata T, Saito A, Mizuno J, Sugimoto H, Noguchi K, Kikuchi E, Inui H
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this study, we have proposed and demonstrated experimentally a novel monitoring device of single mouse embryo activity after in vitro fertilization (IVF) using a semiconductor-based field effect transistor (FET). The FET biosensor realized to detect it noninvasively, quantitatively, and continuously by change of hydrogen ions with positive charges, which were induced by dissolved carbon dioxide due to cellular respiration activity during cleavage. The electrical signal of FET biosensor should become an effective indication to evaluate objectively single embryo activity as its morphology is observed subjectively after IVF. The platform based on the FET biosensor will contribute to promote elective single embryo transfer (eSET) in human assisted reproductive technology (ART).