2013

An electrochemical one-step system for assaying methyltransferase activity based on transport of a quantum dot signaling tracer.

Biosensors & bioelectronics Baek S, Won BY, Park KS, Park HG
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A one-step, electrochemical method for assaying methyltransferase (MTase) activity, based on the convective transport of a quantum dot (QD) signaling tracer, has been developed. The assay chip used in this system was prepared by modifying a gold matrix with CdSe/ZnS QD-tagged dsDNA, which contains a specific methylation site (5'-GATC-3') recognized by MTase. Treatment of the chip with DNA adenine methylation (Dam) MTase, generates a methylated sequence (5'-GAmTC-3') within the dsDNA. The methylated dsDNA is then subjected to a cleavage reaction, induced by DpnI, which leads to release from the gold matrix of a DNA fragment tethered to a QD. Detection of the released QD, using square wave anodic stripping voltammetry (SWASV) on a glassy carbon (GC) electrode, enables the reliable quantitation of the methylated DNA. Because it is accomplished in a simple and convenient one step and does not require any complicated secondary or tedious washing steps, the new assay method holds great promise for epigenetic analysis in facility-limited environments or point-of-care testing (POCT) applications.

关键词