2013

Aptamer-linked biosensor for thrombin based on AuNPs/thionine-graphene nanocomposite.

The Analyst Zhang Z, Luo L, Zhu L, Ding Y, Deng D, Wang Z
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A novel protocol of electrochemical thrombin aptasensor based on a goldnanoparticles/thionine-graphene (AuNPs/Th-G) nanocomposite modified glassy carbon electrode was presented. Graphene was non-covalently functionalized by thionine viaπ-stacking interaction, and then AuNPs were electrodeposited onto the Th-G surface. The morphology and conductivity of the AuNPs/Th-G nanocomposite were characterized by scanning electron microscopy, cyclic voltammetry and electrochemical impedance spectroscopy. The ethanethiol-substituted oligonucleotide probe was immobilized onto the surface of the AuNPs/Th-G nanocomposite to prepare an electrochemical biosensing platform. The aptamer-thrombin reaction was monitored by differential pulse voltammetry. Under optimum conditions, the proposed biosensor exhibits high sensitivity and a low detection limit for thrombin determination. The thrombin could be quantified in a wide range of 0.5 to 40 nM and with a low detection limit of 0.093 nM (S/N = 3).

关键词