2014

Cysteamine-palladium complex ([Pd(μ-OAc)(ppy)]2, ppy:2-phenylpyridine, PhMe)-modified peroxidase biosensor immobilized on a gold electrode.

Artificial cells, nanomedicine, and biotechnology Akyilmaz E, Günay ME, Asav E, Gümüşada R
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A new peroxidase biosensor was developed using cysteamine-palladium complex-modified gold electrode. The principle of the measurements is based on monitoring increase in the oxidation potential of palladium complex (at + 0.47 V vs Ag/AgCl) using amperometric detection. In the optimization studies of the biosensor, effects of enzyme amount, palladium complex amount, and duration of SAM formation on biosensor responses were investigated to optimize the bioactive layer. The biosensor has a fast response time of less than 10 s to hydrogen peroxide (H2O2), with a linear range of 5.0 × 10(- 6) to 150 × 10(- 6) M and a detection limit of 3.38 × 10(- 6) M.

关键词