2013

Enhanced interpretation of adsorption data generated by liquid chromatography and by modern biosensors.

Journal of chromatography. A Agmo Hernández V, Samuelsson J, Forssén P, Fornstedt T
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this study we demonstrate the importance of proper data processing in adsorption isotherm estimations. This was done by investigating and reprocessing data from five cases on two closely related platforms: liquid chromatography (LC) and biosensors. The previously acquired adsorption data were reevaluated and reprocessed using a three-step numerical procedure: (i) preprocessing of adsorption data, (ii) adsorption data analysis and (iii) final rival model fit. For each case, we will discuss what we really measure and what additional information can be obtained by numerical processing of the data. These cases clearly demonstrate that numerical processing of LC and biosensor data can be used to gain deeper understanding of molecular interactions with adsorption media. This is important because adsorption data, especially from biosensors, is often processed using old and simplified methods.

关键词