2013

Direct analysis of trichloropyridinol in human saliva using an Au nanoparticles-based immunochromatographic test strip for biomonitoring of exposure to chlorpyrifos.

Talanta Zhang W, Tang Y, Du D, Smith J, Timchalk C, Liu D, Lin Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A portable immunochromatographic strip-based biosensor for direct detection of trichloropyridinol (TCP), a specific biomarker of exposure to chlorpyrifos, in human saliva sample is presented. In this approach, a series of immunoreactions was performed on the test strip, where the targeted analytes (TCP) bound to the Au nanoparticles-labeled antibodies on the conjugate pad to form analyte-Au-antibody conjugates, and then free Au-labeled antibodies were captured by TCP-BSA in the test zone. Captured Au nanoparticles, increased with decreased levels of analytes, can be observed visibly without any equipment and later quantified by a colorimetric reader. Several experimental parameters were optimized including Au nanoparticle-to-TCP antibody coupling ratio, the amount of Au-labeled TCP antibody, immunoreaction time, the pretreatment of sample pad and the preparation of stock solution of Au-TCP antibody that realize sensitivity, selectivity and direct detection of TCP. Under optimal conditions, this biosensor displays a highly linear range of 0.625-20 ng/mL TCP, with a detection limit of 0.47 ng/mL. Moreover, the immunosensor was successfully used for direct analysis of human saliva sample without any pretreatment. These results demonstrate that this Au nanoparticles-based immunochromatographic test strip (ITS) provides a simple, accurate, and quantitative tool for TCP detection and holds a great promise for point-of-care and in-field analysis of other biomarkers.

关键词