2013

Fabrication of nitric oxide-releasing porous polyurethane membranes-coated needle-type implantable glucose biosensors.

Analytical chemistry Koh A, Lu Y, Schoenfisch MH
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The active release of pharmaceutical agents and the use of porous sensor membranes represent the two most promising strategies for addressing the poor tissue biocompatibility of implantable glucose biosensors. Herein, we describe the combination of these approaches to create nitric oxide (NO)-releasing porous fiber mat-modified sensor membranes. An electrospinning method was used to directly modify needle-type glucose biosensors with the NO donor-loaded fibers. The resulting NO-releasing fiber mat (540 ± 139 nm fiber diameter, 94.1 ± 3.7% porosity) released ~100 nmol of NO per mg of polyurethane over 6 h while maintaining a porous structure without leaching of the NO donor, even in serum. The porous fiber membrane did not influence the analytical performance of the biosensor when ≤50 μm thick.