2013

Measuring chest circumference change during respiration with an electromagnetic biosensor.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Padasdao B, Shahhaidar E, Boric-Lubecke O
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this paper, an off-the-shelf DC motor is modified into a chest belt and used to successfully measure circumference change on a mechanical chest model, while simultaneously harvesting significant power. Chest circumference change can provide information on tidal volume, which is vital in assessing lung function. The chest circumference change is calculated from the motor's voltage output. Calculated values are within 0.95mm of measured circumference changes, with a standard deviation of 0.37mm. The wearable motor can also harvest at least 29.4µW during normal breathing.