2014

Fabrication of a highly sensitive disposable immunosensor based on indium tin oxide substrates for cancer biomarker detection.

Analytical biochemistry Canbaz MÇ, Sezgintürk MK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Anti-HER-3 antibody was used for the first time in a disposable immunosensor based on indium tin oxide (ITO) substrate for HER-3 quantification. Anti-HER-3 was immobilized onto ITO substrate by 3-aminopropyl)triethoxysilane (APTES) and glutaraldehyde. This highly sensitive immunosensor was capable of detecting concentrations of HER-3 down to the femtogram/ml level by investigating changes in the charge transfer resistance (R(ct)) using electrochemical impedance spectroscopy (EIS). Construction of ITO layers was carefully investigated using a broad range of techniques such as voltammetry, EIS, atomic force microscopy (AFM), and scanning electron microscopy (SEM). Meanwhile, in an immunosensor system, the "single frequency impedance" technique was first used for characterization of interaction between HER-3 and anti-HER-3. Eventually, the proposed ITO-based immunosensor was applied to artificial serum samples spiked with HER-3.