Covalent fabrication of methyl parathion hydrolase on gold nanoparticles modified carbon substrates for designing a methyl parathion biosensor.

Biosensors & bioelectronics Liu G, Guo W, Yin Z
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A biosensor based on AuNP modified GC electrodes has been developed for direct detection of methyl parathion. AuNP can be introduced to mixed monolayers of aryldiazonium salt modified GC electrodes by Au-C bonding through aryldiazonium salt chemistry, which provides a stable interface showing efficient electron transfer between biomolecules and electrodes. PEG molecules were introduced to the interface to resist non-specific protein adsorption. AuNP surfaces were further modified with 4-carboxyphenyl followed by covalent immobilization of methyl parathion hydrolase (MPH), a specific biocatalytic enzyme to methyl parathion. Exposure of this interface to methyl parathion resulted in a change in amperometric signal due to the MPH catalytic hydrolysis of methyl parathion producing electroactive compound 4-nitrophenol. This biosensor shows high selectivity, specificity, reproducibility and stability, and is functional for the detection of methyl parathion in real samples. The linear range for detection of methyl parathion is 0.2-100 ppb with a detection limit of 0.07 ppb in 0.1M phosphate buffer at pH 7.0.