2014

Amperometric bienzyme screen-printed biosensor for the determination of leucine.

Analytical and bioanalytical chemistry Labroo P, Cui Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Leucine plays an important role in protein synthesis, brain functions, building muscle mass, and helping the body when it undergoes stress. Here, we report a new amperometric bienzyme screen-printed biosensor for the determination of leucine, by coimmobilizing p-hydroxybenzoate hydroxylase (HBH) and leucine dehydrogenase (LDH) on a screen-printed electrode with NADP(+) and p-hydroxybenzoate as the cofactors. The detection principle of the sensor is that LDH catalyzes the specific dehydrogenation of leucine by using NADP(+) as a cofactor. The product, NADPH, triggers the hydroxylation of p-hydroxybenzoate by HBH in the presence of oxygen to produce 3,4-dihydroxybenzoate, which results in a change in electron concentration at the working carbon electrode, which is detected by the potentiostat. The sensor shows a linear detection range between 10 and 600 μM with a detection limit of 2 μM. The response is reproducible and has a fast measuring time of 5-10 s after the addition of a given concentration of leucine.

关键词