Nanoelectronic impedance detection of target cells.

Biotechnology and bioengineering Esfandyarpour R, Javanmard M, Koochak Z, Harris JS, Davis RW
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Detection of cells is typically performed using optical fluorescence based techniques such as flow cytometry. Here we present the impedance detection of target cells using a nanoelectronic probe we have developed, which we refer to as the nanoneedle biosensor. The nanoneedle consists of a thin film conducting electrode layer at the bottom, an insulative oxide layer above, another conductive electrode layer above, and a protective oxide above. The electrical impedance is measured between the two electrode layers. Cells captured on the surface of the nanoneedle tip results in a decrease in the impedance across the sensing electrodes. The basic mechanisms behind the electrical response of cells in solution under an applied alternating electrical field stems from modulation of the relative permittivity at the interface. In this paper we discuss, the circuit model, the nanofabrication, and the testing and characterization of the sensor. We demonstrate proof of concept for detection of yeast cells with specificity. We envision the sensor presented in this paper to be combined with microfluidic pre-concentration technologies to develop low cost point-of-care diagnostic assays for the clinical setting.

关键词