2014

A high sensitivity MEA probe for measuring real time rat brain glucose flux.

Biosensors & bioelectronics Wei W, Song Y, Shi W, Lin N, Jiang T, Cai X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The mammalian central nervous system (CNS) relies on a constant supply of external glucose for its undisturbed operation. This article presents an implantable Multi-Electrode Array (MEA) probe for brain glucose measurement. The MEA was implemented on Silicon-On-Insulator (SOI) wafer using Micro-Electro-Mechanical-Systems (MEMS) methods. There were 16 platinum recording sites on the probe and enzyme glucose oxidase (GOx) was immobilized on them. The glucose sensitivity of the MEA probe was as high as 489 µA mM(-1) cm(-2). 1,3-Phenylenediamine (mPD) was electropolymerized onto the Pt recording surfaces to prevent larger molecules such as ascorbic acid (AA), 3,4-dihydroxyphenylacetic acid (DOPAC), serotonin (5-HT), and dopamine (DA) from reaching the recording sites surface. The MEA probe was implanted in the anesthetized rat striatum and responded to glucose levels which were altered by intraperitoneal injection of glucose and insulin. After the in vivo experiment, the MEA probe still kept sensitivity to glucose, these suggested that the MEA probe was reliable for glucose monitoring in brain extracellular fluid (ECF).

关键词