组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
N2 incorporated diamond nanowire (N-DNW) film electrochemical biosensor has utilized for the quantitative determination of urea in aqueous solution and urine sample. N-DNW electrode is wet-chemically cleaned (oxidation) by boiling in a mixture of H2SO4 and HNO3 (3:1) at 200°C for 2h to remove graphite. Urease (Urs) and glutamate dehydrogenase (GLDH) are covalently attached to the oxidized N-DNW electrode by activating the COOH group of N-DNW using ethyl(dimethylaminopropyl)carbodiimide as the coupling agent and N-hydroxysuccinimide as activator. Fourier transform infrared spectroscopy and X-ray photoelectron spectroscopy data reveal that carboxylic and hydroxyl functionalized nature of N-DNW electrodes Urs-GLDH immobilized N-DNW (Urs-GLDH/N-DNW) has been successfully utilized in urea biosensor which exhibits good performance in sensitivity (6.18 μA/mg dL/cm(2)), stability (~1 month), reproducibility, lower detection limit (3.87 mg/dL) and fast response time (>10s). Urs-GLDH/N-DNW also exhibits electrochemical response when tested for different concentration of human urine in buffer solution (from 1:9 to 4:6). In addition, Urs-GLDH/N-DNW bioelectrode retains 80% of its initial enzyme activity for <1 month, when stored at 4-6°C in a refrigerator.