组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Glucose sensors with long life span were fabricated and evaluated in 37°C serum to imitate in vivo conditions. Polyaniline nanofibers were electrodeposited on sensor electrodes for excessive glucose oxidase immobilization to extend the lifetime of sensors. The sensitivity of sensors stored in 37°C bovine serum was stable for 20 days without apparent descents, and still remained 45% of its initial value after 44 days. In order to further improve the life span of sensors, a degradable polymer film, poly lactic-co-glycolic acid (PLGA), was used as an out coating layer of glucose sensors. Results showed that the PLGA coated glucose sensors left 80% of its initial sensitivity after 44 days' storage in 37°C bovine serum. The sensors could be potentially used as long-term implantable devices for real time monitoring of glucose levels.