2014

A convenient sandwich assay of thrombin in biological media using nanoparticle-enhanced fluorescence polarization.

Biosensors & bioelectronics Yue Q, Shen T, Wang L, Xu S, Li H, Xue Q, Zhang Y, Gu X, Zhang S, Liu J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A new aptamer biosensor was presented for the detection of thrombin in this work, which was based on fluorescence polarization (FP) using silica nanoparticles as enhancement probe. The silica nanoparticles covered by streptavidin were tagged with a thrombin aptamer (5'-biotin-GGTTGGTGTGGTTGG-3'), which was bound to the surface of silica nanoparticle through the specific interaction between streptavidin and biotin. In the presence of thrombin, it induced the aptamer to form quadruplex structure. When the other thrombin aptamer labeled with fluorescein (5'-FAM-AGTCCGTGGTAGGGCAGGTTGGGGTGACT-3') was added to the above system, a sandwich structure can form at the surface of silica nanoparticles. The fluorescence polarization was therefore enhanced and quantification between fluorescence polarization signal and concentration of thrombin was built. The sensor provided a linear range from 0.6 to 100 nM for thrombin with a detection limit of 0.20 nM (3.29 SB/m, according to the recent recommendation of IUPAC) in a homogeneous media. The same linear range was obtained in spiked human serum samples with a slightly higher detection limit (0.26 nM), demonstrating high anti-interference of the sensor in a complex biological sample matrix. And the sensor can be used to monitor spiked concentration of thrombin level in real human plasma with satisfactory results obtained.

关键词