2014

Disposable amperometric biosensor based on lactate oxidase immobilised on platinum nanoparticle-decorated carbon nanofiber and poly(diallyldimethylammonium chloride) films.

Biosensors & bioelectronics Lamas-Ardisana PJ, Loaiza OA, Añorga L, Jubete E, Borghei M, Ruiz V, Ochoteco E, Cabañero G, Grande HJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A novel biosensor for lactate has been developed, using screen-printed carbon electrodes (SPCE) and lactate oxidase (LOx). The active surface of the electrodes was modified using a dispersion of platinum nanoparticle decorated carbon nanofibers (PtNp-CNF) in poly(diallyldimethylammonium) chloride (PDDA) solution. In this way, sensitive, disposable, low cost and reliable hydrogen peroxide sensors were obtained. The immobilisation of LOx on top of these PtNp-CNF-PDDA/SPCEs resulted in amperometric biosensors with high operational stability. The sensitivity of the optimised lactate biosensor was 36.8 (mA/Mcm(2)) with a linear range of 25-1500 µM. The limit of detection was 11 µM (S/N=3). Reproducibility, selectivity and storage stability were also evaluated. Additionally, the stability of the biosensor was also predicted by a model based on thermal degradation. Finally, lactate in sweat and blood samples was determined in a sport test using LOx/PtNp-CNF-PDDA/SPCEs and commercial biosensors respectively. Based on these data, the validity of the sweat lactate for the determination of the lactate threshold is discussed.