Multianalyte electrochemical biosensor on a monolith electrode by optically scanning the electrical double layer.

Biosensors & bioelectronics Lee SW, Saraf RF
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Redox on an electrode is an interfacial phenomenon that modulates the charge in the electrical double layer (EDL). A novel instrument, the Scanning Electrometer for Electrical Double-layer (SEED) has been developed to measure multiple enzyme reactions on a monolith electrode due to immunospecific binding with a mixture of respective analytes. SEED quantitatively maps the local redox reaction by scanning a laser on the array of enzyme monolayer spots immobilized on the monolith electrode. SEED measures the change in local charge state of the EDL that abruptly changes due to the redox reaction. The measurement spot size defined by the size of the laser beam is ~10 µm. The SEED signal is linearly proportional to the local redox current density and analyte concentration. The specificity is close to 100%. The SEED readout is compatible with microfluidics platform where the signal degrades less than 2% due to the poly(dimethyl siloxane) (PDMS) body.