2014

Titania nanotube-modified screen printed carbon electrodes enhance the sensitivity in the electrochemical detection of proteins.

Bioelectrochemistry (Amsterdam, Netherlands) Mandal SS, Navratna V, Sharma P, Gopal B, Bhattacharyya AJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The use of titania nanotubes (TiO2-NT) as the working electrode provides a substantial improvement in the electrochemical detection of proteins. A biosensor designed using this strategy provided a robust method to detect protein samples at very low concentrations (Cprotein ca 1ng/μl). Reproducible measurements on protein samples at this concentration (Ip,a of 80+1.2μA) could be achieved using a sample volume of ca 30μl. We demonstrate the feasibility of this strategy for the accurate detection of penicillin binding protein, PBP2a, a marker for methicillin resistant Staphylococcus aureus (MRSA). The selectivity and efficiency of this sensor were also validated using other diverse protein preparations such as a recombinant protein tyrosine phosphatase (PTP10D) and bovine serum albumin (BSA). This electrochemical method also presents a substantial improvement in the time taken (few minutes) when compared to conventional enzyme-linked immunosorbent assay (ELISA) protocols. It is envisaged that this sensor could substantially aid in the rapid diagnosis of bacterial infections in resource strapped environments.

关键词