2014

Development of a real-time capacitive biosensor for cyclic cyanotoxic peptides based on Adda-specific antibodies.

Analytica chimica acta Lebogang L, Hedström M, Mattiasson B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The harmful effects of cyanotoxins in surface waters have led to increasing demands for accurate early warning methods. This study proposes a capacitive immunosensor for broad-spectrum detection of the group of toxic cyclic peptides called microcystins (∼80 congeners) at very low concentration levels. The novel analytical platform offers significant advances compared to the existing methods. Monoclonal antibodies (mAbs, clone AD4G2) that recognize a common element of microcystins were used to construct the biosensing layer. Initially, a stable insulating anchor layer for the mAbs was made by electropolymerization of tyramine onto a gold electrode surface, with subsequent incorporation of gold nanoparticles (AuNPs) on the glutaraldehyde (5%) activated polytyramine surface. The biosensor responded linearly to microcystin concentrations from 1×10(-13)M to 1×10(-10)M MC-LR standard with a limit of detection of 2.1×10(-14)M. The stability of the biosensor was evaluated by repeated measurements of the antigen and by determining the capacitance change relative to the original response, which decreased below 90% after the 30th cycle.

关键词