2014

EIS-based biosensor for ultra-sensitive detection of TNF-α from non-diluted human serum.

Biosensors & bioelectronics Kongsuphol P, Ng HH, Pursey JP, Arya SK, Wong CC, Stulz E, Park MK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Serum background is a critical issue for biosensor development as it interferes with the detection of target molecules and may give rise to false positive signal. We present here highly sensitive and selective TNF-α biosensor which is able to detect TNF-α from non-diluted human serum using magnetic bead coupled antibody and electrochemical impedance spectroscopy (EIS) techniques. The process is designed to detect TNF-α from human serum in three stages; (1) abundant protein backgrounds are depleted from the serum using magnetic bead coupled albumin and IgG antibodies, (2) after background depletion TNF-α is captured using magnetic bead coupled TNF-α antibody, and (3) the captured TNF-α is eluted from the magnetic beads and measured using EIS technique in which comb structured gold microelectrodes array (CSGM) is utilized to enhance the detection sensitivity. The system is able to achieve the limit of detection (LOD) at 1 pg/ml (57 fM) and a linear relationship between increasing TNF-α concentrations and charge-transfer resistance in a dynamic range of 1-1000 pg/ml.

关键词