Flow injection analysis biosensor for urea analysis in urine using enzyme thermistor.

Applied biochemistry and biotechnology Mishra GK, Sharma A, Deshpande K, Bhand S
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

There is a need for analytical methods capable of monitoring urea levels in urine for patients under clinical monitoring to appraise renal function. Herein, we present a practical method to quantify levels of urea in human urine samples using flow injection analysis-enzyme thermistor (FIA-ET) biosensor. The biosensor comprises a covalently immobilized enzyme urease (Jack bean) on aminated silica support, which selectively hydrolyzes the urea present in the sample. Under optimized conditions, the developed biosensor showed a linear response in the range of 10-1,000 mM, R (2) = 0.99, and response time of 90 s in 100 mM phosphate buffer (PB) (flow rate of 0.5 mL/min, sample volume of 0.1 mL, and pH 7.2). The urea-spiked human urine samples showed minimal matrix interference in the range of 10-1,000 mM. Recoveries were obtained (92.26-99.80 %) in the spiked urine samples. The reliability and reproducibility of the developed biosensor were found satisfactory with percent relative standard deviation (% RSD) = 0.741. The developed biosensor showed excellent operational stability up to 30 weeks with 20 % loss in original response when used continuously at room temperature. These results indicate that the developed biosensor could be very effective to detect low and high levels of urea in urine samples.