2014

A new method to assay hypoxia-inducible factor-1 based on small molecule binding DNA.

Analytica chimica acta Yang N, Li H, Yin T, Huang Y, Sun L, Li G
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Hypoxia-inducible factor-1 (HIF-1) is among the most important indicators of hypoxia in evaluating severity of many diseases. In this work, a novel method for HIF-1 detection is proposed by using electrochemical techniques based on small molecule binding DNA. In this method, since the designed DNA sequence can specifically bind with either an electroactive small molecule or HIF-1, the signal readout is inversely proportional to HIF-1 concentration, thus a simple and easily-operated method for HIF-1 detection can be developed. With the proposed method, HIF-1 can be determined in a linear range from 5 to 25nM with a detection limit of 2.8nM. Furthermore, the proposed method can be directly used to assay HIF-1 in placenta tissue, and the assay results can reliably reflect the severity of preeclampsia, a very dangerous condition during pregnancy. The proposed method also shows desirable sensitivity, high selectivity and excellent reproducibility, so this method can have potential applications in clinical practice.